Report from BESSY
Monday Oct 20, 1997 thru 4 a.m. local time (2 UT) Oct 21

Report from BESSY, Monday morning thru evening with comments thru Tues. morning

Operator: B. Wargelin

Configuration: SSD_5 on SX700, SSD_X on 12-20 (white beam)

The entire PTB crew was in a meeting that lasted all morning, and did not begin work on setup until 1 p.m. It then took until 7 p.m. for them to complete their setup, which included photodiode beam intensity calibration.

Frank Scholze recommended using the larger 1.2-mm aperture on the SX700 because it provides a more accurate calibration than the 140-um ap. The ap is about 1 m from the Ge crystal, and beam divergence adds another ~2 mm to the spot size. We therefore used the 5-mm SSD aperture for all tests. (Tuesday morning we compared rates with the 2-mm and 5-mm aps. They were the same to within a few percent, indicating the beam size is ~<2 mm when it hits the SSD.)

BESSY no longer runs at 850 MeV, so we ran at 800 MeV.

Beam current is recorded continuously, so no MCB gating is required. For the record, the HXDS computer reads about 10 seconds ahead of PTB's.



SSD_5 on SX700

All msmts are Spectral Response Function/QE. The QE maps are lower priority and were skipped for this first time through. All runs 1000 sec unless noted.

Test runs listed as: RunID, CMDB-ID, Energy, start (UT), and additional notes


The remainder of tests listed below were run on the next shift.  They are listed here as copied from Brad's email report which includes comments thru Tuesday morning, October 21, 4:00 am local.


SSD_X on White light (12-20) beam

All measurements are of relative aperture size.  Integration times are 100 seconds.

Test runs listed as: RunID, CMDB-ID, Aperture Size

For runs 500370-374, ignore the autoDeadTime_pulses. The BH-1 pulser is feeding both SSDs, but the TTL output from the DB-2 pulser was being fed into the deadtime counter. The pulser was very stable, however, so the measurements are still valid. For subsequent runs, the TTL output from the BH-1 feeds both deadtime counters.

The rest of the beam was lost at about 3:30 a.m. Tuesday local time. There was no reinject until 7 a.m. Tuesday morning.